Micromac ECO

Online Analyzer for COD/BOD/pH/TSS or COD/Color/pH/TSS MICROMAC ECO is a microprocessor-controlled online analyzer specifically designed for automatic measurements at the outlet of wastewater treatment plants.

Robust and Reliable: Designed for industrial applications, it guarantees maximum electronic, mechanical, and hydraulic reliability. The complete separation of the electronic and hydraulic systems, combined with the extreme simplicity and robustness of the hydraulic circuit, ensure safe and reliable long-term operation.

Automatic Calibration: At intervals set by the operator, the analyzer performs automatic calibration, triggering an alarm if the measured calibrant value is abnormal.

Automatic calibrationAt times specified by the operator, the analyser performs automatic calibration, producing an alarm signal if the measured calibrant value is anomalous.

Test frequency: Programmable by the operator; between tests, the analyzer enters standby mode without consuming reagents.

Reanalysis of out-of-range results: The analyzer detects out-of-range results and reanalyzes them “in dilution.”

Advantages and Benefits

  • 8-inch color touchscreen;
  • funzionamento completamente automatico;
  • long service life, minimal maintenance, low operating costs;
  • non utilizza reagenti;
  • Ease of use: No special training is required
  • completely separate hydraulic and electronic systems;
  • serial interface for connecting to a PC or printer (optional);
  • Annual maintenance.

Technical Specifications

MEASUREMENT PRINCIPLE UV COD: SAC at 254 nm, turbidity correction at 550 nm according to DIN standard 348404-3; COD correlation with a known calibration sample
BOD: correlation with COD; the BOD–COD correlation can be adjusted and/or established on-site
TSS: turbidity measurement at 850 nm according to ISO 7027, correlation with TSS to be established on-site. Standard flow cell with an optical path length of 2.5 mm; other lengths available upon request: 3.0/5/10 mm; LED lifespan: 5 years
pH: combined pH electrode in the flow cell, temperature-controlled sample, pH values corrected for temperature
STANDARD SERIES COD: 0–50/100/200/500 mg/L – alternative ranges available upon request
BOD: 0–15/25/50/100 mg/L – alternative ranges available upon request
TSS: 0–20/50/100/150 NTU – alternative ranges available upon request
pH: Calibration range 4–7–10
Measurement type Batch, multiparametric batch and sequential version
Measurement frequency Programmable
Measurement time COD/BOD: 7–10 min / TSS: 7–10 min / pH: 7–10 min; depending on the measurement range and sample temperature; for combined versions, 10–15 min
MEASUREMENT POINTS COD: up to 6 streams; BOD: up to 6 streams; COD/BOD: up to 3 streams; COD/BOD/TSS or pH: up to 2 streams; same range across all streams. MP4 version (unit with full options) supports single stream only
OUTPUT SIGNAL AND COMMUNICATION PORTS Bidirectional RS232 – standard, or, upon request, one of the following:
a) RS485 – optional
b) Modbus RTU – optional
c) Separate 4–20 mA outputs for each method or flow – optional
Galvanic isolator on the 4–20 mA outputs – optional
Entrance Signals Analysis: 1 digital contact with photocoupler, galvanically isolated
Calibration: 1 digital contact with photocoupler, galvanically isolated
ALARM CONTACTS Standard:
General: 1 SPDT dry-contact relay, maximum load 24 V AC/DC 0.5 A, separate for each flow - Standard
Options (separate for each method):
Upper limit: 1 potential-free SPDT switch, maximum load 24 V AC/DC 0.5 A
Calibration: 1 potential-free SPDT switch, maximum load 24 V AC/DC 0.5 A
ALARM MESSAGES On an 8” color touchscreen
Sample delivery Pressure: atmospheric
Temperature: 10° - 35 °C
Volume: 50/80 mL per analysis, depending on the analyzer model
Connection: standard 2x4 silicone, other types available upon request
Outlet: 2x4 mm pressureless silicone
REAGENTS Not required; only washing solutions and calibration solutions
Room temperature 10–45°C
ASSEMBLY Wall-mounted
LEVEL OF PROTECTION IP 55, IP 65 optional on request
HARDWARE PC104 standard microcontroller, integrated 8" colour touchscreen
Feeding 12 Vdc; external power supply (110/220 Vac to 12 Vdc) included with the product
ABSORPTION 4W in standby, 10W in analysis
WEIGHT 25 kg without reagents
DIMENSIONS 800x420x280 mm (height x width x depth)
SELF-CLEANING FILTER
Feeding 12 Vcc
SAMPLE PRESSURE min 0.3 bar
SAMPLE FLOW RATE 30 l/h max 1 bar
SELF-CLEANING COMPRESSED AIR Max 2 bar

Documentation

Brochure

Technical and Commercial Information

Select a parameter to download the relevant technical data sheet

FAQs

Lorem ipsum dolor sit amet, consectetur adipiscing elit. Ut elit tellus, luctus nec ullamcorper mattis, pulvinar dapibus leo.

Lorem ipsum dolor sit amet, consectetur adipiscing elit. Ut elit tellus, luctus nec ullamcorper mattis, pulvinar dapibus leo.

Related Products and Models

Micromac 2000 - fully automated analyser

Micromac 2000

EASYCHEM TOX Early Warning - Robotic Technology Direct Reading Analyser

Easychem TOX Early Warning

Easychem Coli Online

Easychem Coli Online

EASYCHEM TOX Early Warning - Robotic Technology Direct Reading Analyser

Easychem Online

Request information

Connect with our experts to find the perfect solution for your needs
your company. Schedule a demo or request a quote today.