The analyzer consists of an industrial cabinet housing the multiparameter instrument platform, which includes a refrigerated compartment for reagents, controls, and calibrators, a rotating mechanical arm for sampling and dispensing reagents and samples, a temperature control system, and a reaction plate equipped with 80 reaction cuvettes, along with a multi-wavelength spectrophotometer and a washing station with a drain.
| Standard | EPA Method 245.7 – Mercury in Water by Cold Vapour Atomic Fluorescence Spectrometry ISO 17852:2006 – Water quality Determination of mercury Method using atomic fluorescence spectrometry HJ 926-2017 – Technical requirements and monitoring methods for water quality with mercury Online automated monitor |
| Analysis method | Cold vapour atomic fluorescence method |
| Measuring range | 0-1/2/5 µg/L (expandable) |
| Precision | ≤ 5% |
| Display value error | ≤ 10% |
| Ionic interference | ≤ 20% |
| Zero-point drift | ≤ 2% |