[vc_row tm_bgimagefixed=””][vc_column][vc_column_text]

SH-UVFS200 UV full spectrum absorption digital probe

Overview

The UV full spectrum absorption digital probe employs UV-Vis absorption technology to ascertain corresponding pollutant levels within water samples, achieved through dual-beam compensation integrated with chemometric algorithmic analysis. The measurement system comprises the measurement probe, controller, connecting cable (standard 10 metres), and a mechanical brush.

Measurement technique

Different pollutants in the water body have certain characteristics of light absorption at different wavelengths (water absorption spectral fingerprint), as shown in the following figure. The full-spectrum water quality probe detects pollutants in the water body using the absorption spectrum in the 190nm ~ 750nm band and a chemometric mapping relationship, so as to obtain the analysis of COD, TOC, BOD5, turbidity, colour, TSS, and nitrate pollutant content in water samples.

Product features

 

Probe technical specifications

 

Measured wavelength 190~750 nm
Path length 35 / 15 / 5 mm, selectable
Measurement parameters COD, TOC, BOD, Nitrate, Turbidity, TSS, Colour, SAC254
Response time Less than 10 seconds
Measurement range (35 mm path length)

COD: (0-70) mg/L; TOC: (0-30) mg/L;

Nitrate: (0-5) mg/L;

Turbidity: (0-200) NTU; TSS: (0-100) mg/L;

Colour: (0~300) Hazen;

SAC254: (0 to 70) Abs/m.

Measurement range (15mm path length)

COD: (0 to 165) mg/L; TOC: (0 to 70) mg/L;

Nitrate: (0–15) mg/L;

Turbidity: (0~465) NTU; TSS: (0~230) mg/L;

Colour: (0-700) Hazen;

SAC254: (0 to 165) Abs/m.

Measurement range (5mm path length)

COD: (0–500) mg/L; TOC: (0–210) mg/L;

Nitrate: (0–50) mg/L;

Turbidity: (0-1400) NTU; TSS: (0-700) mg/L;

Colour: (0–2100) Hazen;

SAC254: (0 to 500) Abs/m.

Performance indicators

 Accuracy: no more than ±5.0% F.S.; Repeatability: ≤2.0%;

Zero drift: no more than ±2% F.S.;

Automatic cleaning The mechanical brush can be set to clean the optical window at regular intervals.
Serial connection RS485
Power supply 24 Vdc / 10 W
Operating temperature 0-50 °C
Protection class IP68
Dimensions 44 (Φ) * 533 (L) mm
Weight around 2.5 kg.

 

Controller technical specifications

User interface 7-inch colour touchscreen
Power supply 220VAC or 24VDC
Method of communication 4G, HJ-212 compliant
Protection class IP65
Operating temperature -20 to 60°C
Dimensions 223 (L) * 199 (L) * 116 (H) mm

 

Technical specifications for mechanical brushes

Cleaning method Automatic cleaning/manual cleaning
Control method Switching signals
Operating temperature (0~50) °C
Power supply 12VDC/1W
Protection class IP68
Dimensions 184 (W) x 85 (D) x 63 (H) mm

 

Flow-cell technical specifications

Internal volume About 100 mL
Inlet/Outlet Pipe diameter Outer diameter 8 mm, inner diameter 5 mm
Materials Main body: polyurethane; viewing window: plexiglass
Dimensions 184 (W) x 85 (D) x 63 (H) mm

 

Product applications

Surface water, sewage plant process sections, storm sewerage network, river outfalls